Section: New Results

Diffusion MRI

A partition of unity finite element method for computational diffusion MRI

D. V. Nguyen, J. Jansson, J. Hoffman and J.-R. Li.

The Bloch-Torrey equation describes the evolution of the spin (usually water proton) magnetization under the influence of applied magnetic field gradients and is commonly used in numerical simulations for diffusion MRI and NMR. Microscopic heterogeneity inside the imaging voxel is modeled by interfaces inside the simulation domain, where a discontinuity in the magnetization across the interfaces is produced via a permeability coefficient on the interfaces. To avoid having to simulate on a computational domain that is the size of an entire imaging voxel, which is often much larger than the scale of the microscopic heterogeneity as well as the mean spin diffusion displacement, smaller representative volumes of the imaging medium can be used as the simulation domain. In this case, the exterior boundaries of a representative volume either must be far away from the initial positions of the spins or suitable boundary conditions must be found to allow the movement of spins across these exterior boundaries.

Many approaches have been taken to solve the Bloch-Torrey equation but an efficient high performance computing framework is still missing. In this paper, we present formulations of the interface as well as the exterior boundary conditions that are computationally efficient and suitable for arbitrary order finite elements and parallelization. In particular, the formulations are based on the partition of unity concept which allows for a discontinuous solution across interfaces conforming with the mesh with weak enforcement of real (in the case of interior interfaces) and artificial (in the case of exterior boundaries) permeability conditions as well as an operator splitting for the exterior boundary conditions. The method is straightforward to implement and it is available in FEniCS for moderate-scale simulations and in FEniCS-HPC for large-scale simulations. The order of accuracy of the resulting method is validated in numerical tests and a good scalability is shown for the parallel implementation. We show that the simulated dMRI signals offer good approximations to reference signals in cases where the latter are available and we performed simulations for a realistic model of a neuron to show that the method can be used for complex geometries.

Diffusion MRI simulation in thin-layer and thin-tube media using a discretization on manifolds

D. V. Nguyen, J. Jansson, H. T. A. Tran, J. Hoffman and J.-R. Li.

The Bloch-Torrey partial differential equation describes the evolution of the transverse magnetization of the imaged sample under the influence of diffusion-encoding magnetic field gradients inside the MRI scanner. The integral of the magnetization inside a voxel gives the simulated diffusion MRI signal. This paper proposes a finite element discretization on manifolds in order to simulate the diffusion MRI signal in domains that have a thin layer or a thin tube geometrical structure. Suppose that the three-dimensional domain has a thin layer structure: points in the domain can be obtained by starting on the two-dimensional manifold and moving along a depth (thickness) function. For this type of domains, we propose a finite element discretization formulated on a surface triangulation of the manifold. The variable thickness of the domain is included in the weak formulation on the surface triangular elements. A simple modification extends the approach to `thin tube' domains where a manifold in one dimension and a two-dimensional variable cross-section describe the points in the domain. We conducted a numerical study of the proposed approach by simulating the diffusion MRI signals from the extracellular space (a thin layer medium) and from neurons (a thin tube medium), comparing the results with the reference signals obtained using a standard three-dimensional finite element discretization. We show good agreement between the simulated signals using our proposed method and the reference signals. The approximation becomes better as the diffusion time increases. The method helps to significantly reduce the required simulation time, computational memory, and difficulties associated with mesh generation, thus opening the possibilities to simulating complicated structures at low cost for a better understanding of diffusion MRI in the brain.

The time-dependent diffusivity in the abdominal ganglion of Aplysia californica, experiments and simulations

K. V. Nguyen, D. Le Bihana, L. Ciobanua and J.-R. Li

The nerve cells of the Aplysia are much larger than mammalian neurons. Using the Aplysia ganglia to study the relationship between the cellular structure and the diffusion MRI signal can potentially shed light on this relationship for more complex organisms. We measured the dMRI signal of chemically-fixed abdominal ganglia of the Aplysia at several diffusion times. At the diffusion times measured and observed at low b-values, the dMRI signal is mono-exponential and can be accurately represented by the parameter ADC (Apparent Diffusion Coefficient).

We performed numerical simulations of water diffusion for three types of cells in the abdominal ganglia: the large cell neurons, the bag cells, and the nerve cells. For the bag cells and nerves cells, we created spherical and cylindrical geometrical configurations that are consistent with known information about the cellular structures from the literature. We used the simulation results to obtain information about the intrinsic diffusion coefficient in these cells.

For the large cell neurons, we created geometrical configurations by segmenting high resolution T2-weighted (T2w) images to obtain the cell outline and then incorporated a manually generated nucleus. We used numerical simulations to validate the claim that water diffusion in the large cell neurons is in the short diffusion time regime for our experimental diffusion times.

Then, using the analytical short time approximation (STA) formula for the ADC, we showed that in order to explain the experimentally observed behavior in the large cell neurons, it is necessary to consider the nucleus and the cytoplasm as two separate diffusion compartments. By using a two compartment STA model, we were able to illustrate the effect of the highly irregular shape of the cell nucleus on the ADC.

The derivation of homogenized diffusion kurtosis models for diffusion MRI

H. Haddar, M. Kchaou and M. Moakher

We use homogenization theory to establish a new macroscopic model for the complex transverse water proton magnetization in a voxel due to diffusion-encoding magnetic field gradient pulses in the case of biological tissue with impermeable membranes. In this model, new higher-order diffusion tensors emerge and offer more information about the structure of the biological tissues. We explicitly solve the macroscopic model to obtain an ordinary differential equation for the diffusion MRI signal that has similar structure as diffusional kurtosis imaging models. We finally present some validating numerical results on synthetic examples showing the accuracy of the model with respect to signals obtained by solving the Bloch-Torrey equation.

On-going collaborative projects on DMRI

J.R. Li, H. Haddar and I. Mekkaoui

  • We performed simulations for a collaborative project with Demian Wassermann of the Parietal team on distinguishing between Spindle and pyramidal neurons with Multi-shell Diffusion MRI.

  • We continue in the simulation and modeling of heart diffusion MRI with the post-doc project of Imen Mekkaoui, funded by Inria-EPFL lab. The project is co-supervised with Jan Hesthaven, Chair of Computational Mathematics and Simulation Science (MCSS), EPFL.