EN FR
EN FR


Section: New Results

Motion Control of Cable-Driven Continuum Catheter Robot through Contacts

Catheter-based intervention plays an important role in minimally invasive surgery. For the closed-loop control of catheter robot through contacts, the loss of contact sensing along the entire catheter might result in task failure. To deal with this problem, we propose a decoupled motion control strategy which allows to control insertion and bending independently. We model the catheter robot and the contacts using the Finite Element Method. Then, we combine the simulated system and the real system for the closed-loop motion control. The control inputs are computed by solving a quadratic programming (QP) problem with a linear complementarity problem (LCP). A simplified method is proposed to solve this optimization problem by converting it into a standard QP problem. Using the proposed strategy, not only the control inputs but also the contact forces along the entire catheter can be computed without using force sensors. Finally, we validate the proposed methods using both simulation and experiments on a cable-driven continuum catheter robot for the real-time motion control through contacts [17].

Figure 7. (a) and (c) present experimental setups for the validation. (b) and (d) show that we take into account the contact and provide accurate results in the simulation
IMG/Catheter.png